PERFORMANCE TESTS OF ACTIVE PERSONAL DOSEMETER DEVELOPED BY ÇNRTC/NEL     
Yazarlar
Prof. Dr. Doğan YAŞAR Prof. Dr. Doğan YAŞAR
Kırşehir Ahi Evran Üniversitesi, Türkiye
Enis Kapdan
Muharrem Korkmaz
Özet
The Active personal dosemeter (APD) developed by Cekmece Nuclear Research and Training Center/Nuclear Electronic Laboratory (CNRTC/NEL) was calibrated at the reference radiation quality (Cs-137). The performance tests were carried out with Cs-137, Co-60 and X-ray beam qualities in compliance with the standards of International Organization for Standardization ISO4037-1,3 and The International Electrotechnical Commission, IEC61526:2010. Response of the personal dose equivalent was observed to be within the range from -7.6% to +4.2%. The dose equivalent rates were observed with the maximum deviation -26.8% for 0.1 mSv h(-1) and within the range from -2.3% to + 10.1% for dose rates over 0.5 mSv h(-1). It is found that the APDs have lower readings up to 50% for 33 keV photons and perform a distribution in the acceptable limits for over 33 keV. Angular dependence was tested at 0 degrees, +/- 30 degrees and +/- 60 degrees in orthogonal plane and the angular dependence of the APDs is within acceptable limits except N40.
Anahtar Kelimeler
Makale Türü Özgün Makale
Makale Alt Türü SSCI, AHCI, SCI, SCI-Exp dergilerinde yayımlanan tam makale
Dergi Adı Radiation Protection Dosimetry
Dergi ISSN 0144-8420
Dergi Tarandığı Indeksler SCI
Makale Dili İngilizce
Basım Tarihi 11-2017
Cilt No 176
Sayı 4
Sayfalar 380 / 387
Doi Numarası 10.1093/rpd/ncx021
Makale Linki http://academic.oup.com/rpd/article/176/4/380/3058925
BM Sürdürülebilir Kalkınma Amaçları
Atıf Sayıları
Google Scholar 1
PERFORMANCE TESTS OF ACTIVE PERSONAL DOSEMETER DEVELOPED BY ÇNRTC/NEL

Paylaş