PERFORMANCE TESTS OF ACTIVE PERSONAL DOSEMETER DEVELOPED BY ÇNRTC/NEL      
Yazarlar
Prof. Dr. Doğan YAŞAR
Kırşehir Ahi Evran Üniversitesi, Türkiye
Enis Kapdan
Muharrem Korkmaz
Özet
The Active personal dosemeter (APD) developed by Cekmece Nuclear Research and Training Center/Nuclear Electronic Laboratory (CNRTC/NEL) was calibrated at the reference radiation quality (Cs-137). The performance tests were carried out with Cs-137, Co-60 and X-ray beam qualities in compliance with the standards of International Organization for Standardization ISO4037-1,3 and The International Electrotechnical Commission, IEC61526:2010. Response of the personal dose equivalent was observed to be within the range from -7.6% to +4.2%. The dose equivalent rates were observed with the maximum deviation -26.8% for 0.1 mSv h(-1) and within the range from -2.3% to + 10.1% for dose rates over 0.5 mSv h(-1). It is found that the APDs have lower readings up to 50% for 33 keV photons and perform a distribution in the acceptable limits for over 33 keV. Angular dependence was tested at 0 degrees, +/- 30 degrees and +/- 60 degrees in orthogonal plane and the angular dependence of the APDs is within acceptable limits except N40.
Anahtar Kelimeler
Makale Türü Özgün Makale
Makale Alt Türü SSCI, AHCI, SCI, SCI-Exp dergilerinde yayımlanan tam makale
Dergi Adı Radiation Protection Dosimetry
Dergi ISSN 0144-8420
Dergi Tarandığı Indeksler SCI
Makale Dili İngilizce
Basım Tarihi 11-2017
Cilt No 176
Sayı 4
Sayfalar 380 / 387
Doi Numarası 10.1093/rpd/ncx021
Makale Linki http://academic.oup.com/rpd/article/176/4/380/3058925
BM Sürdürülebilir Kalkınma Amaçları
Atıf Sayıları
WoS 1
Google Scholar 1
PERFORMANCE TESTS OF ACTIVE PERSONAL DOSEMETER DEVELOPED BY ÇNRTC/NEL

Paylaş