PERFORMANCE TESTS OF ACTIVE PERSONAL DOSEMETER DEVELOPED BY ÇNRTC/NEL
Yazarlar (3)
Enis Kapdan
Muharrem Korkmaz
Makale Türü Özgün Makale (SSCI, AHCI, SCI, SCI-Exp dergilerinde yayınlanan tam makale)
Dergi Adı Radiation Protection Dosimetry (Q3)
Dergi ISSN 0144-8420 Dergi Bilgileri (2017)
Dergi Tarandığı Indeksler SCI
Makale Dili İngilizce Basım Tarihi 11-2017
Kabul Tarihi Yayınlanma Tarihi 01-03-2017
Cilt / Sayı / Sayfa 176 / 4 / 380–387 DOI 10.1093/rpd/ncx021
Makale Linki http://academic.oup.com/rpd/article/176/4/380/3058925
UAK Araştırma Alanları
Özet
The Active personal dosemeter (APD) developed by Cekmece Nuclear Research and Training Center/Nuclear Electronic Laboratory (CNRTC/NEL) was calibrated at the reference radiation quality (Cs-137). The performance tests were carried out with Cs-137, Co-60 and X-ray beam qualities in compliance with the standards of International Organization for Standardization ISO4037-1,3 and The International Electrotechnical Commission, IEC61526:2010. Response of the personal dose equivalent was observed to be within the range from -7.6% to +4.2%. The dose equivalent rates were observed with the maximum deviation -26.8% for 0.1 mSv h(-1) and within the range from -2.3% to + 10.1% for dose rates over 0.5 mSv h(-1). It is found that the APDs have lower readings up to 50% for 33 keV photons and perform a distribution in the acceptable limits for over 33 keV. Angular dependence was tested at 0 degrees, +/- 30 degrees and +/- 60 degrees in orthogonal plane and the angular dependence of the APDs is within acceptable limits except N40.
Anahtar Kelimeler
BM Sürdürülebilir Kalkınma Amaçları
Atıf Sayıları
PERFORMANCE TESTS OF ACTIVE PERSONAL DOSEMETER DEVELOPED BY ÇNRTC/NEL

Paylaş