| Yazarlar (1) |
Prof. Dr. Levent URTEKİN
Kırşehir Ahi Evran Üniversitesi, Türkiye |
| Özet |
| This study evaluates the biomechanical performance of Ti6Al-4 V and 316L stainless steel (SS) screws for femoral neck fracture fixation. The microstructural analysis scanning electron microscopy (SEM) and energy-dispersive X-ray spectroscopy (EDS) was conducted to examine material composition. X-ray diffraction patterns of Ti6Al4 V and AISI 316L materials were added for identifying crystallographic phases. Simulation results revealed significant differences when comparing Ti6Al4V and 316L screw materials. For Ti6Al4V screws, the von Mises stress in the screw body was 204.56 MPa, and the stress above the fracture line was 124.57 MPa. In contrast, for 316L stainless steel screws, the stress within the body increased to 236.33 MPa, while the stress at the fracture line decreased to 109.65 MPa. Examining deformation-related parameters, the gap value for Ti6Al4V screws was 0.065 mm, penetration value ... |
| Anahtar Kelimeler |
| biomechanics | corrosion | FEA | Femoral neck fracture | shear strength | Wolff's law |
| Makale Türü | Özgün Makale |
| Makale Alt Türü | SSCI, AHCI, SCI, SCI-Exp dergilerinde yayınlanan tam makale |
| Dergi Adı | Proceedings of the Institution of Mechanical Engineers, Part E: Journal of Process Mechanical Engineering |
| Dergi ISSN | 0954-4089 Wos Dergi Scopus Dergi |
| Dergi Tarandığı Indeksler | |
| Dergi Grubu | Q2 |
| Makale Dili | İngilizce |
| Basım Tarihi | 07-2025 |
| Sayı | 1 |
| Sayfalar | 2147483647 / 0 |
| Doi Numarası | 10.1177/09544089251390137 |