This study is aimed to validate 90Y dosimetry with MC simulation by comparing the TLD measurements results obtained in 3D liqid phantom (LP) and its monte carlo (MC) simulation. In the first step, tumor imitation was filled with 1900 µCi (70.3 MBq) amount of 99mTc activity in LP. In addition, 24 thermoluminescence dosimetry chips (TLD-100) were placed in 8 different locations in the LP. The same measurement system was created in GEANT4 Tomography Emission Application (GATE version 8.1). DoseActor library was defined for 24 points instead of TLDs and doses were taken. In the second step, 90Y with the amount of 2 mCi (74 MBq) activity and 27 DoseActor libraries were placed in tumor mimic and different locations in LP to calculate the dose distribution with GATE. GATE and TLD results for 99mTc were compared for validation. The strong correlation coefficient was calculated between the instant ... |