img
KFORCE Sens® electrogoniometer in the evaluation of wrist range of motion in patients with carpal tunnel syndrome: Validity and reliability   
Yazarlar
İsmail Ceylan
Şafak Kuzu
Mehmet Canlı
Anıl Özüdoğru
Halil Alkan
Figen Tuncay
Özet
This study aimed to test the validity and reliability of the KFORCE Sens® (electrogoniometer) in carpal tunnel syndrome (CTS) patients. Thirty-one volunteers (aged 32-60 years) diagnosed with unilateral CTS were included in this study. Each of the two raters evaluated the participants' wrist range of motion (ROMs) (flexion, extension, radial deviation, and ulnar deviation) twice, for a total of four times, using an electrogoniometer. In addition, wrist ROMs were evaluated once with a universal goniometer. According to the study findings, both the inter-rater and test-retest reliability of the electrogoniometer device in all ROMs of the wrist were excellent (ICC>0.80). In addition, the ROM values of the electrogoniometer device and the ROM values of the universal goniometer in the same joint were highly correlated (r>0.80, p<0.05). The electrogoniometer is a valid, reliable, and practical device for evaluating wrist ROMs in patients with CTS.
Anahtar Kelimeler
Makale Türü Özgün Makale
Makale Alt Türü Uluslararası alan indekslerindeki dergilerde yayımlanan tam makale
Dergi Adı Journal of Experimental and Clinical Medicine
Dergi Tarandığı Indeksler
Makale Dili İngilizce
Basım Tarihi 10-2024
Cilt No 41
Sayı 3
Sayfalar 505 / 509
BM Sürdürülebilir Kalkınma Amaçları
Atıf Sayıları

Paylaş