img
Computer aided analysis of biomechanical performance of schanz screw with different additive manufacturing materials used in pertrochanteric fixator on an intertrochanteric ...
Yazarlar
Prof. Dr. Levent URTEKİN Prof. Dr. Levent URTEKİN
Kırşehir Ahi Evran Üniversitesi, Türkiye
Özet
This study examines the use of computer‐aided analysis to evaluate the biomechanical performance of Schanz screws made from different additive manufacturing materials (Ti6Al4V, 316 L, Inconel 625, and Inconel 718) in a pertrochanteric fixator for the treatment of intertrochanteric femoral fractures. Intertrochanteric fractures (ITFs) are severe traumas often seen in the elderly population and can lead to serious consequences. The primary objective of ITF surgery is to provide stability and allow for early ambulation and rehabilitation. The Pertrochanteric Fixator is a surgical implant used to treat hip fractures near the greater trochanter, and is attached to the femur with screws. The procedure is performed under general anesthesia and typically takes 1--2 h. Possible complications include infection, nerve injury, and hardware failure. The aim of this study is to evaluate the biomechanical performance of Schanz ...
Anahtar Kelimeler
Makale Türü Özgün Makale
Makale Alt Türü SSCI, AHCI, SCI, SCI-Exp dergilerinde yayımlanan tam makale
Dergi Adı International Journal for Numerical Methods in Biomedical Engineering
Dergi Tarandığı Indeksler
Dergi Grubu Q2
Makale Dili İngilizce
Basım Tarihi 12-2023
Cilt No 39
Sayı 12
BM Sürdürülebilir Kalkınma Amaçları
Atıf Sayıları
Google Scholar 3

Paylaş