Yazarlar (10) |
![]() |
![]() |
![]() |
![]() |
![]() |
![]() Türkiye |
![]() |
![]() |
![]() |
![]() |
Özet |
A gridded ionization chamber used as a focal plane detector at the back of the TAMU-MDM spectrometer was modified to use MicroMegas technology for the purpose of improving energy resolution and particle identification. The upgraded system was tested in experimental conditions with several heavy-ion beams at 12 MeV/u and found to achieve resolutions between 3.2% and 4.8%. This is a significant improvement over the previous performance of 10–15% obtained using the existing, conventional ionization chambers. |
Anahtar Kelimeler |
Makale Türü | Özgün Makale |
Makale Alt Türü | SSCI, AHCI, SCI, SCI-Exp dergilerinde yayınlanan tam makale |
Dergi Adı | Nuclear Instruments and Methods in Physics Research Section B: Beam Interactions with Materials and Atoms |
Dergi ISSN | 0168-583X Wos Dergi Scopus Dergi |
Dergi Tarandığı Indeksler | SCI |
Makale Dili | İngilizce |
Basım Tarihi | 06-2016 |
Cilt No | 376 |
Sayfalar | 364 / 368 |
Doi Numarası | 10.1016/j.nimb.2016.02.011 |
Makale Linki | http://linkinghub.elsevier.com/retrieve/pii/S0168583X16001282 |
Atıf Sayıları | |
Google Scholar | 9 |