Several formulas to calculate the refractive power for intraocular lens (IOL) implantation are available. Various biometers are on the market for this purpose with built-in programs to directly perform the necessary calculations; eg, the IOLMaster (Carl Zeiss Meditec). It is for the surgeon to decide which instrument and what formula to use to obtain the most appropriate IOL power for his or her patient. Comparing the IOL power results of the Hoffer Q formula 1, 2 for an emmetropic, myopic, and hyperopic eye calculated first with the original (corrected2) Hoffer Q formula and then with the IOLMaster versions 4.02 and 5.02, we found deviations from the results as shown in Table 1. No deviations were found when the IOLMaster’s corneal index of refraction (IR) was adjusted to 1.3375 instead of the originally published 1.336. This probably means that the Hoffer Q formula in the IOLMaster 4.02/5.02 uses a corneal index of …